Skip to main content

Table 2 Gamma index reflecting consistency between the deep-seated dose distributions without and with arrays coverage

From: Impact of transducer arrays on deep-seated dosimetry in radiotherapy with concurrent TTFields for glioblastoma (extreme analysis)

Index

Gamma index (10% threshold)

2%/2 mm

2%/3 mm

3%/2 mm

3%/3 mm

#1

97.5%

98.8%

99.1%

99.3%

#2

87.5%

96.4%

94.6%

98.4%

#3

98.2%

99.2%

99.6%

99.7%

#4

83.1%

87.9%

91.5%

94.3%

#5

98.1%

98.7%

98.6%

98.7%

#6

94.5%

96.4%

97.4%

98.3%

#7

95.4%

97.3%

97.7%

98.3%

#8

91%

93.9%

95.7%

96.7%

\(\overline x \)±std

(93.16 ± 5.16)%

(96.08 ± 3.49)%

(96.77 ± 2.54)%

(97.96 ± 1.61)%